Hot biopsy forceps are widely used as surgical instruments for removal of colorectal polyps during therapeutic colonoscopy; however, special care is needed during the treatment process in order to prevent potential tissue damage such as excessive burning and perforation [1]. Studies on in vivo porcine models have proved the general relationship between power output and tissue damages [2], but the use of these in vivo models limited the application of the results. When compared with in vivo testing models, ex vivo testing models have more advantages: universal setup minimizes the variance induced by the individual lab animals, experiment parameters and conditions; test models generate reliable, comparable results, and avoid the cost of lab animals and related ethical issues. One important factor of the ex vivo test models is the emulation of the properties of a human or a lab animal body. For hot biopsy forceps, the...
Skip Nav Destination
Article navigation
September 2013
Technical Briefs
Novel Ex Vivo Testing Model for Monopolar Hot Biopsy Forceps
Liang Chen,
Liang Chen
Department of Biomedical Engineering
Search for other works by this author on:
Ramiro Malgor,
Ramiro Malgor
Department of Biomedical Sciences
Search for other works by this author on:
David Drozek,
David Drozek
Heritage College of Osteopathic Medicine
Search for other works by this author on:
JungHun Choi
JungHun Choi
Department of Mechanical Engineering,
Ohio University
Search for other works by this author on:
Liang Chen
Department of Biomedical Engineering
Ramiro Malgor
Department of Biomedical Sciences
David Drozek
Heritage College of Osteopathic Medicine
JungHun Choi
Department of Mechanical Engineering,
Ohio University
Manuscript received March 15, 2013; final manuscript received April 2, 2013; published online July 3, 2013. Assoc. Editor: Arthur G. Erdman.
J. Med. Devices. Sep 2013, 7(3): 030943 (2 pages)
Published Online: July 3, 2013
Article history
Received:
March 15, 2013
Revision Received:
April 2, 2013
Citation
Chen, L., Malgor, R., Drozek, D., and Choi, J. (July 3, 2013). "Novel Ex Vivo Testing Model for Monopolar Hot Biopsy Forceps." ASME. J. Med. Devices. September 2013; 7(3): 030943. https://doi.org/10.1115/1.4024347
Download citation file:
13
Views
0
Citations
Get Email Alerts
Cited By
Accuracy of an Apparatus for Measuring Glenoid Baseplate Micromotion in Reverse Shoulder Arthroplasty
J. Med. Devices (December 2022)
Three-Dimensional Printing in Combined Cartesian and Curvilinear Coordinates
J. Med. Devices (December 2022)
A Novel Low-Cost Three-Dimensional Printed Brace Design Method for Early Onset Scoliosis
J. Med. Devices (September 2022)
Related Articles
2014 Richard Skalak Award
J Biomech Eng (February,2015)
Endometrial Thermal Balloon Ablation Using a High Temperature, Pulsed System: A Mathematical Model
J Biomech Eng (December,2003)
Thermal Assessment of Ablation Limit of Subsurface Tumor During Focused Ultrasound and Laser Heating
J. Thermal Sci. Eng. Appl (March,2016)
Method for Detecting Ablation Damage Using Sensor Integration of Hot Biopsy Forceps
J. Med. Devices (September,2013)
Related Chapters
Pressure Testing
Process Piping: The Complete Guide to ASME B31.3, Third Edition
Concluding remarks
Mechanical Blood Trauma in Circulatory-Assist Devices
Damage Equivalence of Foam Core Sanswich Composites Subjected to Low-Velocity Impact and Quasi-Static Indentation Testing
International Conference on Advanced Computer Theory and Engineering, 4th (ICACTE 2011)